Dual-energy X-ray absorptiometry (DEXA) is an imaging technique. It provides whole body and regional estimates of the three main body components: fat, lean soft tissues and bone mineral mass. Some software can estimate visceral fat from the android/abdominal region (validated only in adults).
The instruments use a source that generates X-rays at two different energies, a detector, and an interface with a computer system for imaging the scanned areas of interest. The differential attenuation of the X-ray beam at these two energies is used to calculate the bone mineral content (BMC) and soft tissue composition in the scanned region. DEXA uses ionising radiation but the effective dose equivalents of contemporary instrumentation are below or equivalent to background levels, allowing for safe longitudinal studies in both children and adults.
There are two approaches:
The difference in attenuation between the two energy peaks is particular to each element, and thus to each tissue. In all systems, the differential attenuation of the two energies is used to estimate bone mineral content and soft tissue composition.
There are different DEXA technologies:
Calibration procedures
DEXA manufacturers have different quality assurance (QA) and quality control (QC) procedures. These include the system’s calibration as well as testing some software features.
Measurement procedures
Training in DEXA operation, positioning of the participants and data management is essential to obtain reliable data. Training is generally provided by each company along with an electronic standard operating manual embedded in the DEXA software.
The measurement procedure is similar for each instrument:
Figure 1 Showing artefacts: pillow and hip replacement.
Source: MRC Epidemiology Unit.
Figure 2 Body positions for DEXA scan.
Source: MRC Epidemiology Unit.
DEXA estimates bone mineral mass, lean mass and fat mass from the differential absorption of X-rays of two different energies using the instrument's specific algorithms. Manufacturers of DEXA systems also include correction algorithms for the effect of body thickness.
Assumptions associated with DEXA include: the assumed constant attenuation (R) of fat (R = 1.21) and of bone mineral content; minimal effects of hydration on lean tissue estimates.
Setting regions of interest (ROIs)
Different DEXA systems have an automated placement/demarcation of the boundaries of the body regions. Although, readjustments may be required and the standardised protocol in Figure 3 can be adopted.
1: Head: the Head cut is located immediately below the chin.
2: Left and right arm: Both arm cuts pass through the arm sockets and are as close to the body as possible.
3: Left and right forearm: Both forearm cuts are as close to the body as possible and separate the elbows and forearms from the body.
4: Left and right spine: Both spine cuts are as close to the spine as possible without including the rib cage.
5: Left and right pelvis: both pelvis cuts pass through the femoral necks and do not touch the pelvis.
6: Pelvis top: The pelvis top is immediately above the top of the pelvis.
7: Left and right leg: Both leg cuts separate the hands and forearms from the legs.
8: Centre leg: The Centre leg cut separates the right and left leg.
Figure 3 Standard protocol for regions of interest (ROIs).
Source: MRC Epidemiology Unit.
Raw data
DEXA data are stored in databases inbuilt within the DEXA systems such as Microsoft Access and SQL. A data extraction tool can be set up to access the raw data and reconstruct body composition variables where necessary.
The raw data includes estimates of:
Figure 4 Locations of body composition estimations (e.g. whole, android region etc.).
Source: MRC Epidemiology Unit.
Bone mineral density standard scores
DEXA also estimates bone mineral density T-score and Z-score.
T-score
Table 1 World Health Organization cut off values to indicate presence of osteoporosis.
T-score | Interpretation according to WHO criteria |
---|---|
Above -1 | Bone density is considered normal |
Between -1 and -2.5 | Sign of osteopenia, a condition in which bone density is below normal an may lead to osteoporosis |
Below -2.5 | Indicates presence of osteoporosis |
Adapted from: [4].
Z-score
Body composition variables when symmetry method is applied
Some software allows automated estimations of body composition for scans where the symmetry method was applied (e.g. arm omitted from scanning area or half-scan analysis). However, it is preferable to switch this feature off as the estimation can occur even if, for example, small tissue is clipped from the scanning area.
The following steps can be followed to rebuild an individual’s body composition data, depending on which regions are omitted from the scan. After the missing tissue has been rebuilt, percentage body fat can be derived.
a) Calculations for symmetry method when the left arm omitted is from scanning area
The procedure below removes the bone, the fat and the lean masses of the left arm from total mass, left total mass and left arm mass values and replaces it with the right arm values.
b) Calculations for 1/2 body symmetry method
c) Calculation of body composition following rebuild
Percentage body fat can then be derived using the following calculations:
Visceral fat estimates
Different manufacturer have developed algorithms to estimate visceral fat from the android/abdominal region by estimating the subcutaneous fat on each side of the abdominal wall thickness (Figure 5). This measurement is used to extrapolate the amount of subcutaneous fat in the abdominal/android region using appropriate modelling (e.g. circular or spherical models). Visceral fat is derived by subtracting the estimated abdominal subcutaneous fat from the total android/abdominal fat mass.
Figure 5 Visceral fat estimation within the android/abdominal region.
Source: MRC Epidemiology Unit.
Derived variables
Further data can be derived, which are not given by the software automatically:
Harmonisation of data from different DEXA manufacturers
There are published translation regression equations to tackle the systematic differences in the absolute values of body composition variables between different manufactures [2, 9]. However, these equations have been derived from healthy individuals and they might not be generalisable to other populations. Cross-calibration study of different instruments is therefore recommended to derive translation regression equations.
When replacing an instrument, the ISCD (International Society of Clinical Densitometry) guidelines recommend carrying out cross-calibrations between the old and new instruments in at least 30 participants scanned on each instrument, twice on the new instrument to test precision and accuracy and derived cross-calibration equations.
DEXA was created for and continues to be used as a clinical tool to assess bone mineral density and bone mineral content to diagnose osteoporosis (thinner bones). Other body sites (e.g. spine, forearm and hip scans) are used for this purpose.
It is typically used in research settings to assess body composition in:
An overview of the characteristics of whole body DEXA scanning is outlined in Table 3.
Strengths
Table 2 Precision of two DEXA instruments.
Total body fat CV% | Regional body fat CV% | Pixel (mm) | |
---|---|---|---|
Prodigy | 1.9 | 1.2 – 4.4 | 4.8 / 13.0 |
iDEXA | 0.9 | 0.9 – 2.4 | 2.4 / 3.04 |
CV: Coefficient of variation.
Adapted from: [3].
Figure 6 Fat patterning and distribution using DEXA.
Source: MRC Epidemiology Unit.
Limitations
Table 3 Characteristics of whole body DEXA scanning.
Consideration | Comment |
---|---|
Number of participants | Large |
Relative cost | Medium |
Participant burden | Low |
Researcher burden of data collection | Low |
Researcher burden of coding and data analysis | Medium |
Risk of reactivity bias | No |
Risk of recall bias | No |
Risk of social desirability bias | No |
Risk of observer bias | No |
Space required | Medium |
Availability | High |
Suitability for field use | Possible only using mobile van |
Participant literacy required | No |
Cognitively demanding | No |
Considerations relating to the use of DEXA for anthropometry in specific populations are described in Table 4.
Table 4 Anthropometry by whole body DEXA scan in different populations.
Population | Comment |
---|---|
Pregnancy | Not suitable due to the unknown adverse effects of the radiation dose to the fetus. |
Infancy and lactation | Swaddling techniques are recommended in infants and the measurements can be carried out while the infant is asleep. However, the accuracy of regional soft tissue measurements can be affected if swaddling is used as the arm regions will be incorporated with the thigh regions. |
Toddlers and young children | This technique may be impractical in young children as the protocol requires the participant to stay as still as possible as movements may invalidate the results. Feasible to use down to 4 years and in small infants. |
Adolescents | Suitable. |
Adults | Suitable. |
Older Adults | Suitable. |
Ethnic groups | Reference data for body composition not available for some ethnic groups. |
Other (obesity) | Most systems cannot accommodate individuals with BMI ≥35. Newer models can measure up to 200 kg. A half body scan is generally required in those individuals (see measurement procedure). This method provides valid estimates of body composition when compared to the criterion method. |
Safety and ethical considerations:
Other procedural considerations:
Refer to section: practical considerations for objective anthropometry
A method specific instrument library is being developed for this section. In the meantime, please refer to the overall instrument library page by clicking here to open in a new page.